A nonlinear image contrast sharpening approach based on Munsell's scale

Sean C Matz1, Rui J P de Figueiredo

  • 1Boeing Company, Seal Beach, CA 90740, USA.

Related Concept Videos

Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been developed.
Calibration Curves: Linear Least Squares01:20

Calibration Curves: Linear Least Squares

A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
Skewness01:06

Skewness

The measures of central tendency calculated from a data set may not reveal much about its intrinsic distribution. If a plot is made of the data set’s values, the mean and the median may not only differ, but also the plot may have more values on one side of the central tendencies. Such a data set is said to be skewed towards that side.
The longer the tail of the plot on one side, the more skewed it is. The skewness of a data set’s values suggests that the measures of central tendency are...
Dose Response Curve: Conventional Versus Nonmonotonic01:21

Dose Response Curve: Conventional Versus Nonmonotonic

The correlation between a drug's dosage and its impact on a biological system is a cornerstone of pharmacology and toxicology. Conventional dose–response curves, which include graded and quantal relationships, are key to this understanding. Graded dose–response curves depict the spectrum of a biological reaction to different doses within an individual, indicating that as the drug dosage increases, so does the intensity of the response. On the other hand, quantal dose–response relationships...