Related Experiment Video
Updated: Aug 9, 2026

Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Half-life measurement of 55Fe
R Van Ammel1, S Pommé, G Sibbens
1European Commission, Joint Research Centre, Institute for Reference Materials and Measurements, Retieseweg 111, B-2440 Geel, Belgium. raf.van-ammel@cec.eu.int
Abstract:
The half-life of 55Fe was determined experimentally and compared with literature values. A solid 55Fe source was prepared by electrodeposition and repeatedly measured in an X-ray detector at a constant, low solid angle. In total, more than six hundred measurements were performed over a period of ten months, corresponding to one third of the half-life. The experimental method and the corresponding uncertainty budget are presented. The result of 1005.2+/-1.4 days is slightly higher than the current recommended values.