Related Experiment Video
Updated: Aug 9, 2026

3D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
Richardson-Lucy algorithm with total variation regularization for 3D confocal microscope deconvolution
Nicolas Dey1, Laure Blanc-Feraud, Christophe Zimmer
1Ariana Group, INRIA/I3S, 2004 route des Lucioles-BP93, 06902 Sophia Antipolis, France.
Abstract:
Confocal laser scanning microscopy is a powerful and popular technique for 3D imaging of biological specimens. Although confocal microscopy images are much sharper than standard epifluorescence ones, they are still degraded by residual out-of-focus light and by Poisson noise due to photon-limited detection. Several deconvolution methods have been proposed to reduce these degradations, including the Richardson-Lucy iterative algorithm, which computes maximum likelihood estimation adapted to Poisson statistics. As this algorithm tends to amplify noise, regularization constraints based on some prior knowledge on the data have to be applied to stabilize the solution. Here, we propose to combine the Richardson-Lucy algorithm with a regularization constraint based on Total Variation, which suppresses unstable oscillations while preserving object edges. We show on simulated and real images that this constraint improves the deconvolution results as compared with the unregularized Richardson-Lucy algorithm, both visually and quantitatively.
Related Concept Videos
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Three-Dimensional Microscopy in Microbiology