Abstract:
A simple, rapid, in vitro cell-free protein expression system, Expressway NMR, is introduced and used to express the small ubiquitin-related modifier protein SUMO-1. This 12 kDa molecule is challenging for NMR as it has limited solubility and requires relatively high salt (200 mM) for stability in solution. Starting with the gene, the cell-free system, and milligram amounts of nitrogen-15 isotopically enriched amino acids, sufficient protein is produced in 4 h to obtain a high-resolution 2D HSQC spectrum of the protein in 40 min. This time would be closer to 10 min with the aid of a higher sensitivity salt-tolerant cryogenic NMR probe. With all protein purification steps included, and aggressive data processing using the filter diagonalization method (FDM), it is but 6 h from gene to heteronuclear single quantum coherence (HSQC). As the cell-free system is nearly background-free, it is also possible to work with the crude reaction mixture, in which case only a total of 5 h is required. Sample stability over time, whether crude extract or purified, was notable, with no significant change in the 15N-1H HSQC spectrum over 6 months at 4 degrees C (300 muM, pH 6.1, capped NMR tube). The combination of a turnkey, high-yield, protease-free in vitro protein expression system, an optimized sensitivity-enhanced HSQC pulse sequence, and FDM processing makes this scheme an attractive first step to rapidly assess the suitability of proteins for complete solution structure determination.
More Related Videos
Related Concept Videos
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
2D NMR: Overview of Heteronuclear Correlation Techniques