Operating characteristics of the Finapress system to predict elevated left ventricular filling pressure
C Gillard1, C Henuzet, J Lallemand
1Service of Cardiology, CHU A. Vesale, Montigny-le-Tilleul, Belgium.
Background:
Dyspnea is a common symptom and can be caused by many different conditions. The detection of congestive heart failure (CHF) is sometimes difficult.
Hypothesis:
The pulse amplitude ratio (PAR) measured by the Finapress procedure during a Valsalva maneuver can detect elevated left ventricular end-diastolic pressure (LVEDP) accurately over a wide range of values.
Methods:
Comparison of the estimated LVEDP by PAR with the invasively measured LVEDP before and after ventriculography during coronography was made in 101 consecutive stable patients referred for chest pain and/or chronic dyspnea.
Results:
A significant correlation was found between the catheter-measured LVEDP (range 3-40 mmHg) and the PAR (R2 = 0.70, p < 0.001). The receiver operator characteristics (ROC) of the PAR to detect an LVEDP > 15 mmHg can be considered to be excellent, with an area under the ROC curve achieving 0.92 (95% confidence interval [CI] 0.87-0.96; p < 0.001). A PAR of > 0.675 predicted the presence of an LVEDP > 15 mmHg with a sensitivity of 0.865 (95% CI 0.780-0.926) and a specificity of 0.847 (95% CI 0.730-0.928). The positive and negative LRs were 5.70 and 0.16, respectively.
Conclusions:
The observed likelihood ratios confirm that the PAR determined by the Finapress procedure may be a useful bedside diagnostic tool in patients with cardiac conditions.