Fluorescence quencher improves SCANSYSTEM for rapid bacterial detection
M Schmidt1, M K Hourfar, A Wahl
1Institute of Transfusion Medicine and Immunohematology, German Red Cross, Johann Wolfgang Goethe University, Frankfurt, Germany. mschmidt@bsdhessen.de
Background And Objectives:
The optimized scansystem could detect contaminated platelet products within 24 h. However, the system's sensitivity was reduced by a high fluorescence background even in sterile samples, which led to the necessity of a well-trained staff for confirmation of microscope results.
Materials And Methods:
A new protocol of the optimized scansystem with the addition of a fluorescence quencher was evaluated. Pool platelet concentrates contaminated with five transfusion-relevant bacterial strains were tested in a blind study.
Results:
In conjunction with new analysis software, the new quenching dye was able to reduce significantly unspecific background fluorescence. Sensitivity was best for Bacillus cereus and Escherichia coli (3 CFU/ml).
Discussion:
The application of a fluorescence quencher enables automated discrimination of positive and negative test results in 60% of all analysed samples.