Related Experiment Video
Updated: Aug 9, 2026

A Model of Long-Term Ventricular Fibrillation in Isolated Rat Hearts
Published on: February 17, 2023
The effect of induction method on defibrillation threshold and ventricular fibrillation cycle length
Endre Zima1, Mihály Gergely, Pál Soós
1Cardiovascular Center, Department of Cardiovascular Surgery, Semmelweis University, 68 Városmajor Str., H-1122 Budapest, Hungary.
Introduction:
Since no clinical data are available on the comparison of the "shock on T-wave" and "high frequency burst" ventricular fibrillation (VF) induction modes during defibrillation threshold (DFT) testing, we aimed to compare these two methods during implantable cardioverter defibrillator implantation.
Methods:
The DFT was determined with a step-down protocol using biphasic, anodal polarity (100%, 40%, 20% voltage control) shocks. Patients were randomized: VF was induced by 50 Hz burst in group B (n = 45) and T-wave shock in group T (n = 41). The DFT was defined as the lowest energy level that terminated VF; confirmed DFT (DFTc) was defined as the minimal energy level that consecutively terminated VF twice. Success rate of DFTc was calculated during an intraindividual test for the alternate induction method.
Results:
A total of 546 episodes of VF were induced: n = 278 (B) vs n = 268 (T). Incidence of VT during inductions was 9.9% (B) vs 2.7% (T), P < 0.05. Neither the DFT, 8.8 +/- 4.0 J (B) vs 9.7 +/- 4.2 J (T), nor the DFTc, 10.6 +/- 5.1 J (B) vs 10.8 +/- 4.2 J (T), proved to be significantly different. A significant correlation was found between VF cycle length (CL) and the concomitant DFT (r = 0.298, P < 0.05) in group T only. Subgroup analysis of patients under chronic class III antiarrhythmic treatment showed no increase of the DFT in either group and significantly lower incidence of VT induction in group T regardless of antiarrhythmic treatment.
Conclusion:
The DFT and the VFCL proved to be independent of the VF induction method. The T-wave shock was more unlikely to induce VT during DFT testing. These results suggest that both methods are reliable in DFT determination, though T-wave shock application is a more reliable method for DFT testing.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Induction
A...
Electrophysiology of Normal Cardiac Rhythm