Related Experiment Video
Updated: Aug 8, 2026

Characterizing Far-infrared Laser Emissions and the Measurement of Their Frequencies
Published on: December 18, 2015
Absolute measurement of F2-laser power at 157 nm
Stefan Kück1, Friedhelm Brandt, Hans-Albert Kremling
1Physikalisch-Technische Bundesanstalt, Braunschweig, Germany.
Abstract:
We report a comparison of laser power measurements at the F2-laser wavelength of 157 nm made at two facilities of the Physikalisch-Technische Bundesanstalt (PTB), the German national metrology institute. At the PTB laboratory at the electron storage ring BESSY II in Berlin, the scale for laser power was directly traced to a cryogenic radiometer operating at 157 nm, whereas at the PTB laser radiometry facility in Braunschweig the calibration of transfer detectors was performed with a newly developed standard for laser power at 157 nm, which is traceable in several steps to a cryogenic radiometer operating at 633 nm. The comparison was performed under vacuum conditions with laser pulse energies of approximately 10 microJ, however with different average powers because different primary standard radiometers were used. The relative deviation for the responsivity of the transfer detector was 4.8% and thus within the combined standard uncertainty.