Related Experiment Video
Updated: Jul 28, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Development of a rotating X-ray shutter for coronary angiography using synchrotron radiation
Abstract:
The first clinical examination using a two-dimensional imaging system of coronary angiography with monochromated synchrotron radiation was carried out at the National Laboratory for High Energy Physics in May 1996. A rotating X-ray shutter was developed to produce a pulsed X-ray beam with 2-6 ms of beam spill to suppress image blurring, at a frequency of 30 Hz. A performance test of the X-ray shutter using synchrotron radiation was carried out, and it was verified that the shutter had satisfactory specifications for clinical applications. With this X-ray shutter the monitored radiation dose in clinical examinations was consistent with theoretical expectations and kept within a reasonable level of radiation protection.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...