Related Experiment Video
Updated: Aug 8, 2026

Cooling an Optically Trapped Ultracold Fermi Gas by Periodical Driving
Published on: March 30, 2017
Determination of cooling rates in a quadrupole ion trap
David M Black1, Anne H Payne1, Gary L Glish2
1Kenan Laboratory, Department of Chemistry, University of North Carolina, 27599-3290, Chapel Hill, NC, USA.
Abstract:
Collisional cooling rates of infrared excited ions are measured in a quadrupole ion trap (QIT) mass spectrometer at different combinations of temperature and pressure. Measurements are carried out by monitoring fragmentation efficiency of leucine enkephalin as a function of irradiation time by an infrared laser after a short excitation and incrementally increasing cooling periods. Cooling rates are observed to be directly related to bath gas pressure and inversely related to bath gas temperature. The cooling rate at typical ion trap operating pressure (1 mTorr) and temperature (room T) is faster than can be measured. At elevated temperature and the lowest pressure used for the studies, the rate of collisional cooling becomes negligible compared to the rate of radiative cooling.
Related Concept Videos
Mass Analyzers: Common Types
Mass Analyzers: Overview
Chemical Ionization (CI) Mass Spectrometry