A Novel Technique for Accurate Intensity Calibration of Area X-ray Detectors at Almost Arbitrary Energy
Abstract:
A novel intensity uniformity calibration method for area X-ray detectors is described. In diffraction experiments, amorphous lithium glass plates, containing doping elements chosen for their K edges just below the energy of the main beam, replace the crystallographic samples for the calibration measurement. The fluorescent emission excited by the X-ray beam is almost isotropic. It has exactly the same geometry as the diffracted radiation, and can be obtained at the same wavelength by proper selection of the element and excitation energy. A sample 2theta scan allows the emission distribution as a function of angle to be characterized with an accuracy of a fraction of a percent. This allows a flat-field correction of similar accuracy. The quality of crystallographic data collected with an X-ray image intensifier/CCD detector was significantly improved by flat-field correction using an Sr-doped lithium tetraborate glass. This technique can be applied to X-ray energies from 5 to 50 KeV; the calibration sample is small, stable and easily handled.
Related Concept Videos
Determination of Crystal Structures
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
X-ray Imaging
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...