Validation of a non-mercury digital auscultatory device with manual pressure registration (PMS Mandaus)
Anja Wilton1, Husain Shabeeh, Clare Cuckson
1Kings College London, Guy's, Kings and St Thomas' School of Medicine, Maternal and Fetal Health Research Unit, Division of Reproductive Health, Endocrinology and Development, St Thomas' Hospital, London, UK.
Objective:
To assess the clinical accuracy of a non-mercury digital auscultatory device (PMS Mandaus) with manual pressure registration in an adult population.
Methods:
The accuracy of the device was assessed by predefined criteria (European Society of Hypertension protocol) in 33 study participants recruited from among patients and staff in a teaching hospital. A series of same-arm sequential blood pressure measurements were taken: first, two observers taking simultaneous mercury readings, followed by a reading with the device. A total of seven readings were taken from each participant in the sitting position. The data were then analysed according to the European Society of Hypertension protocol and the criteria of the Association for the Advancement of Medical Instrumentation.
Results:
The device fulfilled the criteria of the European Society of Hypertension protocol, achieving a pass for both systolic and diastolic pressure. The device also passed the Association for the Advancement of Medical Instrumentation standard (the mean to be within 5+/-8 mmHg) with the mean differences being -3.2+/-3.8 mmHg for systolic blood pressure and -1.8+/-2.9 mmHg for diastolic blood pressure.
Conclusion:
The device performed in a satisfactory manner according to the European Society of Hypertension and the Association for the Advancement of Medical Instrumentation criteria. The systematic under-recording related to the delay of manual pressure registration is not sufficient to preclude clinical use.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: