Simulated clinical evaluation of four fluid warming devices*
M Turner1, I Hodzovic, W W Mapleson
1Anaesthesia and Intensive Care, Royal Gwent Hospital, Newport, UK. matthewwturner@hotmail.com
Laboratory evaluation of four fluid warming devices (Bair Hugger, Hotline, Standard Ranger, Fluido) showed none reached 37°C at room temperature. However, all devices were predicted to reach target temperatures at manufacturer-specified flow rates.
Area of Science:
- Medical Devices
- Biomedical Engineering
- Anesthesiology
Background:
- Maintaining normothermia during medical procedures is crucial for patient safety.
- Various fluid warming devices are used to prevent hypothermia.
- Understanding the performance of these devices under different conditions is essential.
Purpose of the Study:
- To evaluate the fluid warming capabilities of four commonly used devices: Bair Hugger, Hotline, Standard Ranger, and Fluido.
- To compare device performance under gravity flow and pressurized conditions.
Main Methods:
- Laboratory evaluation of fluid warming devices.
- Testing under gravity flow (1m fluid head) and pressurized flow (300 mmHg).
- Measurement of fluid flow rates (70-450 ml/min) and delivered fluid temperatures at 22°C room temperature.
Main Results:
- No device consistently warmed fluid to the target 37°C at tested flow rates and room temperature.
- Standard Ranger (gravity) and Fluido (gravity and pressurized) achieved the highest temperatures (35°C).
- Bair Hugger and Hotline (gravity and pressurized) and Standard Ranger (pressurized) achieved lower temperatures (24-31°C).
Conclusions:
- Under tested laboratory conditions, none of the evaluated fluid warmers achieved 37°C.
- Calculations suggest all devices could reach target temperatures at their respective manufacturer-specified flow rates.
- Device performance is flow-dependent, highlighting the importance of adhering to manufacturer guidelines for optimal warming.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...