Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Determination of quality parameters from statistical analysis of routine TLD dosimetry data
U German1, M Weinstein, O Pelled
1Nuclear Research Centre Negev, P.O. Box 9100, 84190 Beer Sheva, Israel. gur118@inter.net.il
Abstract:
Following the as low as reasonably achievable (ALARA) practice, there is a need to measure very low doses, of the same order of magnitude as the natural background, and the limits of detection of the dosimetry systems. The different contributions of the background signals to the total zero dose reading of thermoluminescence dosemeter (TLD) cards were analysed by using the common basic definitions of statistical indicators: the critical level (L(C)), the detection limit (L(D)) and the determination limit (L(Q)). These key statistical parameters for the system operated at NRC-Negev were quantified, based on the history of readings of the calibration cards in use. The electronic noise seems to play a minor role, but the reading of the Teflon coating (without the presence of a TLD crystal) gave a significant contribution.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Dosage Regimens: Partial Pharmacokinetic Parameters
Data Validation
Key parameters for method validation include:
Dosage Regimen Designs: Nomograms and Tabulations
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Dose Size and Dosing Frequency: Determination Methods