Internal barium shielding to minimize fetal irradiation in spiral chest CT: a phantom simulation experiment
David K Yousefzadeh1, Matthew B Ward, Chester Reft
1Department of Radiology, University of Chicago, 5841 S Maryland Ave, MC 2026, Chicago, IL 60637, USA. dyousefz@midway.uchicago.edu
Purpose:
To use a phantom to prospectively examine the attenuating effect of barium sulfate as an internal shield to protect the fetus.
Materials And Methods:
In an adult-size phantom, 1- and 2-cm-thick acrylic slabs containing 315 or 630 mL of water, 2% or 40% barium sulfate suspension, and a 1-mm lead sheet were placed under the diaphragm. In 17 experiments, fetal dose was measured by using thermoluminescent dosimeters that were placed immediately under (near field) and 10 cm below (far field) the water slab (eight experiments), barium sulfate slab (eight experiments), and lead sheet (one experiment). In a pulmonary embolism protocol, the phantom was scanned with single-detector spiral computed tomography (CT) at 130 kVp and 230 mAs.
Results:
The control radiation dose was 3.60 mSv+/-0.54 (standard deviation) with the water slab at near field, where the uterus dome is at near term, and 0.507 mSv+/-0.07 with the water slab at far field, the uterus position during early gestation. Scattered radiation was attenuated 13% and 21% with 2% barium sulfate and 87% and 96% with 40% barium sulfate, as calculated in the near and far fields, respectively, and 99% with the 1-mm lead sheet. The extrapolated attenuations for 5%-40% barium sulfate suspensions indicated that beyond a 30% suspension, attenuation increased further only slightly.
Conclusion:
Study results in the phantom experiment suggest that fetal irradiation during maternal chest CT can be reduced substantially with barium shielding.