Numerical analysis of the inner flow field of a biocentrifugal blood pump
Related Experiment Video
Updated: Aug 8, 2026

Modeling and Experimental Analysis of the Single-Shaft Coaxial Motor-Pump Assembly in Electrohydrostatic Actuators
Published on: June 13, 2022
Leok Poh Chua1, Guoliang Song, Tau Meng Lim
1School of Mechanical and Production Engineering, Nanyang Technological University, Singapore. mlpchua@ntu.edu.sg
Abstract:
Implantable ventricular assist devices have been regarded as a promising instrument in the clinical treatment of patients with severe heart failures. In this article, a three-dimensional model of the Kyoto-NTN magnetically suspended centrifugal blood pump was generated and a computational fluid dynamics solution of the inner flow field of the pump including the static pressure distributions, velocity profiles, and the shear stress distributions of the blood was presented. The results revealed that reverse flow generally occurred in the impeller blade channels during the operation of the pump, due to the imbalance of the flow and the pressure gradient generated in the blade channels. The flow pattern at the exit of the blade channels was varying with its angular positions in the pump. The reverse flow at the exit of the impeller blade channels was found to be closely related with the static pressure distribution in the volute passage. Higher pressure in the volute caused severe backflow from the volute into the blade channels. To clarify the effects of a moving impeller on the blood, shear stresses of the blood in the pump were investigated according to the simulation results. The studies indicated that at the beginning of the splitter plate and the cutwater, the highest shear stress exceeded 700 Pa. At other regions such as the inlet and outlet of the impeller blade channels and some regions in the volute passage, shear stresses were found to be about 200 Pa. These areas are believed to have a high possibility of rendering blood trauma.
Related Concept Videos
Applications of Integration to Find Blood Flow
Steady, Laminar Flow in Circular Tubes
Couette Flow
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines. However, the...
Application of the Linear Momentum Equation
The goal is to determine the force components in the x and y directions to hold the pipe in place. Since...