Related Experiment Video
Updated: Aug 8, 2026

A Simple Dewar/Cryostat for Thermally Equilibrating Samples at Known Temperatures for Accurate Cryogenic Luminescence Measurements
Published on: July 19, 2016
A system to irradiate and measure luminescence at low temperatures
M W Blair1, E G Yukihara, S W S McKeever
1Oklahoma State University, Physics Department, 145 PS II, Stillwater, OK 74078, USA. mblair@lanl.gov
Abstract:
We have developed a system to irradiate samples and record radioluminescence (RL), optically stimulated luminescence (OSL), and thermoluminescence (TL) at temperatures ranging from -150 degrees C to 200 degrees C. The system consists of a cryostat, an irradiation/stimulation unit fitted with an X-ray tube (40 kV Moxtek) and a quartz window for optical stimulation, and a detection unit that utilises a photomultiplier tube and an interchangeable filter pack. Experiments have been conducted with quartz and albite (a feldspar). TL and OSL experiments show that several optically sensitive trapping states are stable below -50 degrees C. In addition, an increase in OSL is seen as the OSL stimulation temperature is lowered below -50 degrees C, and an increase in RL is apparent as the temperature is lowered during irradiation. This indicates that not only are optically sensitive low temperature traps present but that luminescence becomes more efficient at low temperatures.
Related Concept Videos
Photoluminescence: Applications
Fluorescence and Phosphorescence: Instrumentation
Flame Photometry: Overview