Related Experiment Video
Updated: Aug 8, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Monte Carlo calculation of the ionization chamber response to 60Co beam using PENELOPE
Chul-Young Yi1, Suck-Ho Hah, Min Sun Yeom
1Ionizing Radiation Group, Korea Research Institute of Standards and Science, P.O. Box 102, Yuseong, Taejon 305-600, Korea. cyyi@kriss.re.kr
Abstract:
The calculation of the ionization chamber response is one of key factors to develop a primary standard of air kerma. Using Monte Carlo code PENELOPE, we simulated the cavity response of the plane parallel ionization chamber to the monoenergetic 60Co beam incident normally on a flat surface of the chamber. Two simulation techniques, namely, the uniform interaction technique and the reentrance technique, were introduced. The effect of the input parameters such as C1 (average angular deflection in a single step between hard elastic events), C2 (maximum average fractional energy loss in a step), S(max) (maximum step length) and W(cc) (the lower energy of secondary electrons created as a result of a hard collision) on the simulated cavity dose was evaluated. We found that the simulated cavity response of the graphite and solid air chambers obtained by PENELOPE to the monoenergetic 60Co beam could be consistent with the value expected from the cross-sections of PENELOPE to within 0.2% (one standard deviation) when W(cc) and S(max) were selected carefully.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016