Evaluation of a modified formalism for precision electron beam dosimetry compared to TG-51 and TRS-398
F Okky Agassy1, Jong In Park2, Chul-Young Yi2
1Korea Research Institute of Standards and Science (KRISS), 267 Gajeong-ro, Yuseong-gu, Daejeon 34113, Republic of Korea; University Science and Technology (UST), 217 Gajeong-ro, Yuseong-gu, Daejeon 34113, Republic of Korea; Research Center for Safety, Metrology, and Nuclear Quality Technology - National Research and Innovation Agency (BRIN), Jakarta 10340, Indonesia.
Purpose:
This study aims to evaluate a modified formalism for electron beam reference dosimetry and to compare the results with the cases of previous protocols TG-51 and TRS-398.
Methods:
Measurements were performed using three types of cylindrical chambers (TW30013, FC65-G, and A1SL) and plane-parallel chambers (PPC) (Roos, Advanced Markus, and NACP-02) with electron beams of energies 6, 12, and 18 MeV. Absorbed dose to water at the reference depths was determined following the modified formalism using different chambers and methods, including the direct application of 60Co calibration or cross-calibration for PPCs in a high-energy electron beam. Data were also analyzed following TG-51 and TRS-398 formalisms for comparison.
Results:
Following the modified formalism, all the results obtained using cylindrical chambers, 60Co calibrated well-guarded PPCs and cross-calibrated PPCs were in good agreement within the estimated uncertainty (0.7 - 0.9 %) at all the energies. However, one PPC exhibited strange behavior, producing 3 % higher doses when applying 60Co calibration. This was confirmed due to chamber-to-chamber variation in Pwall. The modified formalism showed good agreement with TG-51 and TRS-398, but yielded consistently higher doses. The difference was reduced by applying the updated EPOM (effective point of measurement) shift to the previous protocols.
Conclusions:
This study supports the reliability of the modified formalism for electron beam reference dosimetry with chambers calibrated under 60Co, thereby reducing uncertainty. The modified formalism yielded consistent results within smaller uncertainty (0.7 - 0.9 %). However, it is recommended to check each PPC's chamber-to-chamber variation.