Related Experiment Video
Updated: Aug 8, 2026

An Automated Microscopic Scoring Method for the γ-H2AX Foci Assay in Human Peripheral Blood Lymphocytes
Published on: December 25, 2021
Radiation exposure in body computed tomography examinations of trauma patients
M Kortesniemi1, T Kiljunen, A Kangasmäki
1HUS Helsinki Medical Imaging Center, University of Helsinki, PO Box 340, FI-00029 HUS, Finland. mika.kortesniemi@hus.fi
Abstract:
Multi-slice CT provides an efficient imaging modality for trauma imaging. The purpose of this study was to provide absorbed and effective dose data from CT taking into account the patient size and compare such doses with the standard CT dose quantities based on standard geometry. The CT examination data from abdominal and thoracic scan series were collected from 36 trauma patients. The CTDI(vol), DLP(w) and effective dose were determined, and the influence of patient size was applied as a correction factor to calculated doses. The patient size was estimated from the patient weight as the effective radius based on the analysis from the axial images of abdominal and thoracic regions. The calculated mean CTDI(vol), DLP(w) and effective dose were 15.2 mGy, 431 mGy cm and 6.5 mSv for the thorax scan, and 18.5 mGy, 893 mGy cm and 14.8 mSv for the abdomen scan, respectively. The doses in the thorax and abdomen scans taking the patient size into account were 34% and 9% larger than the standard dose quantities, respectively. The use of patient size in dose estimation is recommended in order to provide realistic data for evaluation of the radiation exposure in CT, especially for paediatric patients and smaller adults.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET