Ganfeng Wang1, Yunsheng Hsieh, Xiaoming Cui
1Department of Drug Metabolism and Pharmacokinetics, Schering-Plough Research Institute, 2015 Galloping Hill Road, Kenilworth, NJ 07033, USA. ganfeng.wang@spcorp.com
Abstract:
This paper describes the development and partial validation of a fast, sensitive and specific ultra-performance liquid chromatography/tandem mass spectrometric (UPLC/MS/MS) method for the determination of testosterone (T) and its four metabolites, 6beta-OH-T, 16alpha-OH-T, 16beta-OH-T and 2alpha-OH-T, in in vitro samples. The analytical method involves direct dilution of samples with acetonitrile containing an internal standard, followed by separation of testosterone and the four metabolites on an Acquity UPLCtrade mark C(18) column and detected by selected reaction monitoring (SRM) in positive ionization mode using turbo ionspray ionization. The parent compound and its metabolites investigated were well separated (Rs >1.5) with a run time of 4 min under a gradient condition. The method was partially validated. The linear concentration range was 0.01 to 5 microM for all the compounds of interest. Inter-assay mean bias and relative standard deviation (RSD) were in the range of -12% to 8% and 4.1% to 8.5%, respectively. Intra-assay mean bias and RSD were in the range of -8.0% to 5.2% and 3.4% to 9.6%, respectively. The lower limit of quantitation for this assay was 0.01 microM. The differences in LC/MS performance were investigated by conducting a comparison of UPLC with another method previously optimized for HPLC-based separation and quantification of testosterone and its metabolites.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
05:55Stable Isotope In-Vivo Labeling for Mass-Spectrometry Identification of Paternal Metabolites Transferred from Sperm to Oocyte During Fertilization
Published on: June 17, 2025
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Types of Detectors