Related Experiment Video
Updated: Jul 30, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
The Thompson hemiarthroplasty: a new problem for an old favourite
1Salisbury District Hospital, Salisbury, Wiltshire SP2 8QX, UK. jelloyd1@btinternet.com
Abstract:
The Thompson hemiarthroplasty is a popular hip prosthesis. We present two case reports highlighting a significant alteration in the design of the implant which compromised the success of the operations. In recent years the manufacturing process of this prosthesis has changed, with a resultant increase in the volume of the stem of 10 ml. It is essential that manufacturers inform orthopaedic surgeons of any alteration in the design of the implant and supply compatible instrumentation to minimise surgical errors. Surgeons must remain vigilant when checking the compatibility of the trial and definitive prostheses.