Related Experiment Video
Updated: Aug 7, 2026

A Rat Carotid Artery Pressure-Controlled Segmental Balloon Injury with Periadventitial Therapeutic Application
Published on: July 9, 2020
Thermoelastic analysis of high-pressure angioplasty balloons
J Eaton-Evans1, J M Dulieu-Barton, E Little
1Mechanical and Aeronautical Department, University of Limerick, Ireland.
Abstract:
Failures in angioplasty balloons are investigated using typical destructive techniques. The material properties of moulded balloons are derived from tensile tests and used to establish the reasons for failure of the balloons. Thermoelastic stress analysis is used to determine the stress distribution in the balloons, and a means of interpreting the data to derive actual stresses is described. The departure from linear elastic behaviour in the angioplasty balloons is identified using thermoelastic analysis. The results from the thermoelastic analysis are discussed and compared with those from the destructive tests, and the thermoelastic technique is shown to be a potential new means for non-destructive analysis of angioplasty balloons.
Related Concept Videos
Related Rates
Members Made of Elastoplastic Material
As the bending moment...