[Simultaneous determination of trace platinum, palladium and rhodium in environmental dust samples by ICP-MS after
Yan-zhi Shi1, Run-juan He, Juan Wang
1Center of Analysis and Testing, Capital Normal University, Beijing, China.
Abstract:
A sensitive method for simultaneous determination of Pt, Pd and Rh in dust samples by ICP-MS after open-digestion was developed. The digestion method for the samples and conditions for measurement for ICP-MS were studied in detail. Aquaregia was proved to be a better solvent for the digestion of dust samples. Under the optimum conditions, the limits of detection of Pt, Pd and Rh were 0.48, 4.22 and 0.42 ng x g(-1), respectively. The proposed method was applied to the real sample analysis, and good recoveries were obtained. The recoveries of Pt, Pd and Rh were found to be between 87.6%-88%, 85.2%-87.8%, and 93%-95%, respectively.
More Related Videos
06:53Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
08:46Dose Uptake of Platinum- and Ruthenium-based Compound Exposure in Zebrafish by Inductively Coupled Plasma Mass Spectrometry with Broader Applications
Published on: April 21, 2022
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...