Effect of an active abdominal pulse generator on defibrillation thresholds with a dual-coil, transvenous ICD lead
Eric J Rashba1, Morteza Farasat, Malcolm M Kirk
1Division of Cardiology, University of Maryland School of Medicine, Baltimore, Maryland 21201-1595, USA. erashba@medicine.umaryland.edu
Introduction:
Many patients with implantable cardioverter defibrillators (ICDs) have older lead systems, which are usually not replaced at the time of pulse generator replacement unless a malfunction is noted. Therefore, optimization of defibrillation with these lead systems is clinically important. The objective of this prospective study was to determine if an active abdominal pulse generator (Can) affects chronic defibrillation thresholds (DFTs) with a dual-coil, transvenous ICD lead system.
Methods And Results:
The study population consisted of 39 patients who presented for routine abdominal pulse generator replacement. Each patient underwent two assessments of DFT using a step-down protocol, with the order of testing randomized. The distal right ventricular (RV) coil was the anode for the first phase of the biphasic shocks. The proximal superior vena cava (SVC) coil was the cathode for the Lead Alone configuration (RV --> SVC). For the Active Can configuration, the SVC coil and Can were connected electrically as the cathode (RV --> SVC + Can). The Active Can configuration was associated with a significant decrease in shock impedance (39.5 +/- 5.8 Omega vs. 50.0 +/- 7.6 Omega, P < 0.01) and a significant increase in peak current (8.3 +/- 2.6 A vs. 7.2 +/- 2.4 A, P < 0.01). There was no significant difference in DFT energy (9.0 +/- 4.6 J vs. 9.8 +/- 5.2 J) or leading edge voltage (319 +/- 86 V vs. 315 +/- 83 V). An adequate safety margin for defibrillation (> or =10 J) was present in all patients with both shocking configurations.
Conclusion:
DFTs are similar with the Active Can and Lead Alone configurations when a dual-coil, transvenous lead is used with a left abdominal pulse generator. Since most commercially available ICDs are only available with an active can, our data support the use of an active can device with this lead system for patients who present for routine pulse generator replacement.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...