Related Experiment Video
Updated: Aug 7, 2026

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Utility of the surface ECG before VDD pacemaker implantation
Haran Burri1, Stéphane Noble, Henri Sunthorn
1Cardiology Service, University Hospital, 23, Micheli-du-Crest, 1211 Geneva, Switzerland. haran.burri@hcuge.ch
Surface ECG P-wave amplitude can predict atrial electrogram amplitude in patients receiving VDD pacemakers. A higher P-wave amplitude suggests a lower risk of undersensing, identifying suitable candidates for VDD pacemaker implantation.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Ventricular Demand pacing (VDD) pacemakers are used for atrioventricular block with preserved sinus function, offering a single-lead advantage over DDD systems.
- A significant challenge with VDD pacemakers is the potential for atrial undersensing due to low-amplitude atrial electrograms (EGMs) from the free-floating atrial electrode.
- Currently, there is no reliable method to predict atrial EGM amplitude before implantation, hindering the selection of appropriate candidates for VDD pacing.
Purpose of the Study:
- To determine if the P-wave amplitude measured on a standard surface electrocardiogram (ECG) correlates with the atrial EGM amplitude recorded by a single-pass lead during VDD pacemaker implantation.
- To identify potential predictors of adequate atrial electrogram amplitude for VDD pacemaker recipients.
Main Methods:
- A retrospective review of data from 122 patients who received a VDD pacemaker was conducted.
- Atrial EGM amplitudes measured at the time of implantation were correlated with the maximal P-wave amplitude observed on the surface ECG recorded immediately prior to the procedure.
Main Results:
- A statistically significant positive correlation was found between maximal surface P-wave amplitude and atrial EGM amplitude (Pearson's r=0.313, P<0.001).
- Multivariate analysis confirmed that maximal P-wave amplitude was an independent predictor of atrial EGM amplitude (p=0.003).
- A low atrial EGM amplitude (<0.9 mV) was infrequent overall (9%) but occurred more often in patients with a surface P-wave amplitude ≤0.1 mV (14%) compared to those with >0.1 mV (2%).
Conclusions:
- Low atrial EGM amplitudes requiring potential pacemaker reprogramming are uncommon in patients with single-pass leads for VDD pacing.
- A maximal surface P-wave amplitude greater than 0.1 mV is a strong indicator that a patient is unlikely to experience low atrial EGM amplitude.
- Patients exhibiting a P-wave amplitude >0.1 mV on surface ECG are favorable candidates for VDD pacemaker implantation due to a lower risk of atrial undersensing.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias