Aluminium in peritoneal dialysis fluids as determined by stabilised temperature platform furnace atomic absorption
1Royal Danish School of Pharmacy, Department of Chemistry AD, 2 Universitetsparken, DK-2100 Copenhagen, Denmark.
Abstract:
A procedure is described for the determination of aluminium in peritoneal dialysis fluids. It is based on Zeeman corrected graphite furnace atomic absorption spectrometry employing stabilised temperature platform furnace (STPF) conditions. The samples are analysed after dilution 1 + 1 without pretreatment. The limit of detection of the procedure is 1.5 mug l(-1), and the accuracy and precision were found acceptable. A screening survey showed that aluminium contamination of peritoneal dialysis fluids is not a general problem, but that exceptions to this do occur.
Related Concept Videos
Peritoneal Dialysis I: Introduction and Procedure
Flame Photometry: Lab
Qualitative Analysis
For instance, group IV...