A prototype table-top inverse-geometry volumetric CT system.

Taly Gilat Schmidt1, Josh Star-Lack, N Robert Bennett

  • 1Department of Radiology, Stanford University, Stanford, California 94305, USA. taly.gilat-schmidt@marquette.edu

Medical Physics
|July 29, 2006
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A Position Statement on Endovascular Models and Effectiveness Metrics for Mechanical Thrombectomy Navigation, on Behalf of the Stakeholder Taskforce for Artificial Intelligence-Assisted Robotic Thrombectomy (START).

Journal of the American Heart Association·2026
Same author

OpenDosimeter: Open hardware personal X-ray dosimeter.

Communications engineering·2025
Same author

A virtual trial evaluation of patient motion in arc and linear system designs in body digital tomosynthesis.

Biomedical physics & engineering express·2025
Same author

Analytical model for pulse pileup spectra and count statistics in photon counting detectors with seminonparalyzable behavior.

Medical physics·2025
Same author

Real-World and Clinical Trial Validation of a Deep Learning Radiomic Biomarker for PD-(L)1 Immune Checkpoint Inhibitor Response in Advanced Non-Small Cell Lung Cancer.

JCO clinical cancer informatics·2024
Same author

CT-augmented digital tomosynthesis image reconstruction in image-guided bronchoscopy interventions.

Medical physics·2024
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

A new table-top inverse-geometry CT (IGCT) scanner images a 5-cm volume in one scan without cone-beam artifacts. This volumetric CT system achieves 0.25-mm isotropic resolution, outperforming micro CT for complex objects.

Area of Science:

  • Medical Imaging
  • Computed Tomography

Background:

  • Cone-beam artifacts limit volumetric CT imaging quality.
  • Traditional CT geometries struggle with artifact-free volumetric data acquisition in a single rotation.

Purpose of the Study:

  • To implement and evaluate a table-top inverse-geometry CT (IGCT) scanner.
  • To assess the IGCT system's performance in terms of resolution, noise, and artifact reduction.

Main Methods:

  • Developed a prototype IGCT scanner using a scanned x-ray source and a smaller detector array.
  • Performed scans on a Defrise phantom and inner ear specimens.
  • Compared IGCT performance against a micro CT system and clinical CT.

Main Results:

  • Achieved 0.25-mm isotropic resolution with noise comparable to clinical scanners.

Related Experiment Videos

  • IGCT system faithfully reconstructed phantom details without cone-beam artifacts.
  • Inner ear imaging revealed superior high-resolution details compared to clinical CT.
  • Conclusions:

    • The implemented table-top IGCT system successfully performs single-rotation volumetric scanning.
    • The IGCT geometry effectively eliminates cone-beam artifacts, enabling high-quality volumetric imaging.
    • This technology shows promise for advanced medical imaging applications.