Related Experiment Video
Updated: Aug 6, 2026

Implantation of the Syncardia Total Artificial Heart
Published on: July 18, 2014
Replacement of the left-side valves of an implanted total artificial heart
Keiji Kamohara1, Stephan Weber, Ryan S Klatte
1Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA.
Abstract:
The MagScrew total artificial heart (TAH) is under development. Despite its anticipated durability and reliability, the possibility of a bioprosthetic valve malfunction exists. As a result, the potential for valve replacement surgery, instead of device replacement, would be desirable after a TAH implant. In two of our 90-day animal experiments, we successfully replaced the left-side valves through a left thoracotomy opposite to the right-sided incision site for the initial TAH implant. The results of these cases suggest that the left-side valves could also be replaced through a left thoracotomy approach in humans. To confirm the ability to access the left-side valves in humans, four human cadaver studies were performed with the use of a mock pump designed for human application. This report describes the operative techniques for left-side valve replacement in animals and discusses the advantages of a left thoracotomy in clinical situations, based on results from the human cadaver studies.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Cardiomyopathy V: Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization