Parathyroid adenoma localization: surgeon-performed ultrasound versus sestamibi
David L Steward1, Gregory P Danielson, Chad E Afman
1Department of Otolaryngology-Head and Neck Surgery, University of Cincinnati, Cincinnati, OH 45267, USA. David.Steward@uc.edu
Objectives:
Compare surgeon-performed ultrasound versus sestamibi for preoperative parathyroid adenoma localization.
Study Design:
Single-institutional cohort.
Methods:
One hundred six consecutive patients undergoing parathyroidectomy at an academic institution between 2004 to 2005 were included. Of those, 103 underwent both surgeon-performed ultrasound and sestamibi-Tc99m localization preoperatively. Primary outcome is sensitivity for adenoma localization to correct quadrant (right vs. left, superior vs. inferior).
Results:
Hypercalcemia resolved in 97% of patients. Sensitivities for correct quadrant localization for ultrasound versus sestamibi were 87% versus 58% (P < .001). Specificities were 95%. Positive and negative predictive values were 85% versus 78% and 96% versus 87%, respectively. Combined sensitivity was 93%. Sensitivities for correct side localization were 91% and 74% (P = .002).
Conclusions:
Ultrasound appears more sensitive than sestamibi for localization to correct quadrant or side when performed in-office by the author in this cohort.