Radiation dosimetry of the FluoroCAT scan for real-time endoscopic sinus surgery
Casey R A Manarey1, Vijay K Anand
1Department of Otorhinolaryngology-Head and Neck Surgery, New York Presbyterian Hospital-Weill Cornell Center, Joan and Sanford Weill Medical College of Cornell University, 205 E. 64th Street, New York, NY 10021, USA. caseymanarey@hotmail.com
Objective/Hypothesis:
The FluoroCAT platform of the GE InstaTrak 3500 Plus reconstructs images acquired from the GE OEC 9800 Plus fluoroscope into triplanar images which allow the surgeon to perform real-time endoscopic sinus and skull-base surgery. The radiation dosimetry for the new scanning protocol required by the FluoroCAT has not been reported. This study was designed to measure the radiation dose of a FluoroCAT scan and compare this to the dosimetry of an InstaTrak CT scan.
Study Design:
Prospective study.
Methods:
A standardized CT head phantom was used along with a CT ion chamber to measure radiation exposure. FluoroCAT scans were performed using the GE OEC 9800 Plus fluoroscope. Scans were performed using each of the three fluoroscopic scanning modes on the GE OEC 9800 Plus; normal fluoro, high-level fluoro, and Digital Ciné at 15 pulses per second (pps).
Results:
The highest radiation dose was measured using Digital Ciné 15 pps. The dose measured at the center of the head phantom was 4.1 Gy while the maximum surface dose was 10.7 mGy. The center and surface radiation dose recorded from the CTDI(vol) reading on the control panel of a GE Lightspeed Plus during an InstaTrak CT scan was 85 mGy.
Conclusions:
Using the GE OEC 9800 Plus fluoroscope to accumulate images for the FluoroCAT platform in the InstaTrak 3500 Plus resulted in less radiation compared to an InstaTrak CT scan. This will allow multiple scans to be performed intraoperatively, giving the surgeon the opportunity to perform real-time endoscopic sinus and skull-base surgery.
Ebm Rating:
C-4.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
06:37Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
Published on: October 20, 2010
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):