Cardiac output measurements in off-pump coronary surgery: comparison between NICO and the Swan-Ganz catheter
G Gueret1, G Kiss, B Rossignol
1University Hospital, Department of Anesthesiology and Surgical Intensive Care Unit, Brest, France. gildas.gueret@chu-brest.fr
Background:
The aim of this prospective study was to compare continuous cardiac output measurements of the non-invasive cardiac output system (NICO) with the pulmonary artery catheter during off-pump coronary bypass surgery.
Methods:
Twenty-two patients enrolled for off-pump coronary surgery received both a pulmonary artery catheter and a non-invasive cardiac output system for measurement of cardiac output. Data were compared by the Bland-Altman method to calculate the degree of agreement and to analyse if a significant difference existed between the two methods of cardiac output measurements.
Results:
Perioperatively, the non-invasive cardiac output underestimated cardiac output, but postoperatively overestimated it. The limits of agreement were larger during surgery compared to the postoperative period (-3.1; +2.5 vs. -1.4; +2.2 L min(-1)). Perioperatively, cardiac output measured with the pulmonary artery catheter varied from 0.5 to 7.5 L min(-1) (mean 3.6 L min(-1)) and with the non-invasive cardiac output from 0.5 to 8.4 L min(-1) (mean 3.9 L min(-1)). Postoperatively, these were 2.5-7.7 L min(-1) (mean 4.5 L min(-1)) and 2.3-8.4 L min(-1) (mean 4.9 L min(-1)), respectively.
Conclusion:
During off-pump cardiac surgery, the non-invasive cardiac output reliably measures cardiac output and does it more rapidly than a pulmonary artery catheter and may be more useful in order to detect rapid haemodynamic changes.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization