Related Experiment Video
Updated: Jul 20, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
A separated primary and scatter model for independent dose calculation of intensity modulated radiotherapy
Colin R Baker1, Richard Clements, Alan Gately
1Division of Medical Imaging and Radiotherapy, School of Health Sciences, University of Liverpool, Liverpool, UK. colin.baker@liv.ac.uk
Background And Purpose:
Independent checking of beam monitor units is an essential step in the preparation of any radiotherapy plan. The present work describes a simple model for calculating a point dose from an arbitrary number of irregular, asymmetric fields, typical of beam segments used for IMRT.
Materials And Methods:
Primary and scatter dose contributions were separated using a two-parameter exponential fit to measured beam data, from which differential scatter is determined. A total of 60 IMRT patient plans for a five-field prostate class solution were investigated to validate the model.
Results:
The average difference between the model's prediction and direct measurement of reference dose was found to be -0.6% (ranging from -2.9% to +1.6%), with a standard deviation of 1.0%. This compares well with the observed average difference between treatment planning system prediction and direct measurement of +0.8% (SD 0.6%).
Conclusions:
The model is shown to provide a reliable and accurate independent check of planning system monitor units for the prostate IMRT plans studied. Implementation of the model could significantly reduce the time needed for point dose verification of IMRT plans currently performed by direct measurement.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Biological Effects of Radiation
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses