Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Measurement of energy dependence for XRCT radiochromic film
Martin J Butson1, Tsang Cheung, Peter K N Yu
1Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong, Hong Kong.
Abstract:
Gafchromic XRCT, radiochromic film is assessed over a broad energy range, from kilovoltage to megavoltage x rays for variations in reflected optical density to dose response. A large energy dependence was found with reflected optical density output for the same delivered dose varying from 7.8 +/- 0.35 at 25.5 keV (50 kVp) peaking at 12.1 +/- 0.5 at 54 keV (125 kVp) to 0.975 +/- 0.03 at 2300 keV (10 MV) when normalized to 1 at 1400 keV (6 MV) energy. The response is constant (within 3%) in the 36-69 keV equivalent photon energy range, which corresponds to x-ray tube generating potentials of approximately 100-150 kVp. This matches well with beam qualities for diagnostic computed topography applications.