Related Experiment Video
Updated: Jul 20, 2026

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Evaluation of new image processing conditions for digital mammograms from Fuji computed radiography
Hiroshi Kano1, Tokiko Endo, Mitsuru Ikeda
1Department of Radiology, National Hospital Organization, Nagoya Medical Center, Japan. h.kano@nnh.hosp.go.jp
Purpose:
A new processing parameter (T-type parameter) setting for gray scale was developed to improve the image quality of digital mammograms. To clarify the usefulness of this parameter setting, we have evaluated the image quality of digital mammograms (hard copy images) processed with this parameter, and compared it with S/F mammography.
Material And Methods:
Mammograms were made under the same radiographic conditions by the S/F and FCR systems (type 1, S/F; type 2 and 3, FCR with new T-type parameters; type 4 and 5, FCR with conventional parameters). A total of 49 images from 10 cases was selected for evaluation testing. Evaluation items were the contrast visibility of mammary glands and adipose tissues together with their granularity and sharpness. Eleven medical doctors participated in evaluating the images.
Results:
The FCR hard copy images processed with the T-type parameter settings were significantly preferred over the conventional S/F images for the contrast visibility of mammary glands and adipose tissue. As for the other items (except for granularity), the FCR hard copy images processed with the T-type parameter settings were subjectively evaluated as slightly better than or equal to the S/F images. In contrast, the conventional S/F images were significantly preferred over the FCR hard copy images processed with the conventional parameter settings.
Conclusion:
The image quality of FCR hard copy images processed with the T-type parameter settings was preferred over that of conventional S/F images as evaluated by medical doctors who specialized in mammography interpretation.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Radiological Investigation I: X-ray and CT