Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Comparison of metabolic monitors in critically ill, ventilated patients
Pierre Singer1, Ira Pogrebetsky, Joelle Attal-Singer
1Department of General Intensive Care, Rabin Medical Center, Beilinson Campus, Petah Tiqva, Israel. psinger@clalit.org.il
Objective:
We compared the Deltatrac II, the M-COVX, and the Evita 4 metabolic monitoring devices under clinical conditions.
Methods:
A prospective simultaneous clinical comparison was performed in a general intensive care department of a tertiary university hospital in 43 ventilated, critically ill patients. The monitors were compared simultaneously. After 30 min of steady state, oxygen consumption per unit time, carbon dioxide consumption per unit time, resting energy expenditure, and respiratory quotient were recorded for the Deltatrac II; the same parameters in addition to end-tidal carbon dioxide and fraction of inspired oxygen were recorded for the M-COVX; and carbon dioxide consumption per unit time, end-tidal carbon dioxide, and fraction of inspired oxygen were recorded for the Evita 4. Pulmonary gas-exchange measurements from the Deltatrac II and resting energy expenditure and respiratory quotient from the M-COVX were obtained after 30 min. The other parameters were calculated from the last five measurements obtained at the end of the study period.
Results:
A good correlation was found between oxygen consumption per unit time and resting energy expenditure as obtained from the Deltatrac II and the M-COVX (r = 0.76 and 0.75, respectively; P < 0.001), but the correlation was lower between carbon dioxide consumption per unit time as obtained from the Deltatrac II and the M-COVX or Evita 4 (r = 0.67 and 0.48, respectively). Agreement between the different methods did not reach clinical acceptability, exceeding a 20% difference using the Bland-Altman statistical methods.
Conclusion:
Poor agreement was found between the Deltatrac II and M-COVX or Evita 4 metabolic monitors, despite a good correlation between measurements, leading to the conclusion that the M-COVX and Evita 4 provide less accurate measurements of metabolic gas exchange in stable ventilated patients. These devices can be used for daily nutritional assessment and continuous monitoring, but the Deltatrac II remains the method of choice for metabolic measurement.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation I: Indication and Settings
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.