A constant gradient unilateral magnet for near-surface MRI profiling
Andrew E Marble1, Igor V Mastikhin, Bruce G Colpitts
1MRI Centre, Department of Physics, PO Box 4400, University of New Brunswick, Fredericton, New Brunswick, Canada E3B 5A3.
Abstract:
The design and construction of a unilateral NMR (UMR) magnet assembly for near-surface 1D profiling is presented. The arrangement consists of a single permanent magnet topped with a shaped iron pole cap. The analytically determined profile of the pole cap shapes the field over the magnet, giving a constant gradient of 31 G/cm over a 8mm depth at a 1H frequency of 4.26 MHz in a spot approximately 5 mm wide. The moderate gradient allows 1D profiling of planar samples with a frequency encoded spin-echo experiment. The curvature of the magnetic field limits the available resolution to 100's of microm. The device is suitable for profiling planar samples in which a coarse resolution but large spatial extent is desired.
Related Concept Videos
Magnetic Resonance Imaging
Potential Due to a Magnetized Object
The vector...
Atomic Nuclei: Magnetic Resonance
Magnetic Vector Potential
Consider an ideal solenoid with n turns per unit length and radius R. If I is the current through the solenoid, the magnetic field inside the solenoid is expressed as the product of vacuum...
Magnetic Field Of A Current Loop
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...