Related Experiment Video
Updated: Jul 19, 2026

Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Arterial pulse wave analysis: An accurate means of determining cardiac output in children
Jeffrey J Kim1, W Jeffrey Dreyer, Anthony C Chang
1Section of Pediatric Cardiology, Department of Pediatrics, Baylor College of Medicine, Texas Children's Hospital, Houston, TX, USA.
Insights
Arterial pulse wave analysis (PulseCO) accurately measures cardiac output in children, offering a minimally invasive alternative. This validation study confirms PulseCO
Area of Science:
- Pediatric cardiology
- Hemodynamics
- Medical device validation
Background:
- Cardiac output is crucial for assessing myocardial performance.
- Current methods for cardiac output measurement in children carry risks and challenges.
- Arterial pulse wave analysis (PulseCO) presents a novel, minimally invasive option for continuous cardiac output monitoring.
Purpose of the Study:
- To validate the accuracy of the PulseCO system for noninvasive, real-time cardiac output determination in pediatric patients.
- To assess the clinical agreement between PulseCO and the established thermodilution method in children.
Main Methods:
- Prospective, single-center study involving children with structurally normal hearts undergoing cardiac catheterization.
- Cardiac output measured using the thermodilution technique via a pulmonary arterial catheter.
- Continuous cardiac output data from the PulseCO system were compared with thermodilution results.
Main Results:
- Twenty pediatric patients (median age 10.5 years) were evaluated.
- High correlation (r=0.94, p < .001) observed between PulseCO and thermodilution cardiac index.
- Bland-Altman analysis demonstrated excellent clinical agreement (mean difference 0.19 L/min/m, precision 0.28 L/min/m at 2 sd).
Conclusions:
- The PulseCO system provides a validated, accurate, and minimally invasive method for real-time cardiac output measurement in children.
- PulseCO offers a promising alternative to traditional methods, enhancing patient safety and monitoring convenience.
- This technique supports improved hemodynamic assessment in pediatric cardiology.
Objective:
Cardiac output is a useful measure of myocardial performance. Standard methods of determining cardiac output are not without risk and can be problematic in children. Arterial pulse wave analysis (PulseCO), a novel, minimally invasive cardiac output determination technique, offers the advantage of continuous monitoring, convenience, and low risk. This technique has not been validated in children. The purpose of this study was to validate PulseCO as an accurate means of noninvasively determining real-time cardiac output in children.
Design:
Prospective, single-center evaluation.
Setting:
Children's hospital.
Patients:
Any child with a structurally normal heart, undergoing hemodynamic evaluation in the cardiac catheterization laboratory, was included.
Interventions:
A prograde right heart catheterization was performed, and cardiac output was determined using the thermodilution technique, via placement of a pulmonary arterial catheter.
Measurements And Main Results:
Thermodilution results were compared with continuous real-time cardiac output measurements obtained with the PulseCO system, and they were then analyzed by standard correlation techniques and Bland-Altman analysis. Twenty patients were evaluated with a median age of 10.5 yrs and a median weight of 25 kg. The mean thermodilution cardiac index was 3.3 +/- 0.9 L/min/m, whereas the mean PulseCO cardiac index was 3.1 +/- 0.9 L/min/m. Standard Pearson correlation tests revealed a correlation coefficient of .94 (p < .001). Bland-Altman analysis revealed excellent clinical agreement with a mean difference of 0.19 L/min/m and a precision of 0.28 L/min/m at 2 sd.
Conclusions:
Arterial pulse wave analysis by the PulseCO system provides a novel, minimally invasive method of determining real-time cardiac output in children.
Related Concept Videos
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Special considerations while measuring pulse
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pulse Assessment Sites
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation