Related Experiment Videos

Endocardial pacing in infants and children 15 kg or less in weight: medium-term follow-up

J A Till1, S Jones, E Rowland

  • 1Department of Paediatric Cardiology, Royal Brompton and National Heart Hospital, London, United Kingdom.

Insights

Pediatric pacemaker implantation via transvenous lead placement is generally safe, with most children developing normally. Complications occurred in some, but a novel lead technique showed promise.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering

Background:

  • Transvenous lead placement is a common method for permanent pacemaker implantation in children.
  • Complications associated with pediatric pacemakers can impact patient outcomes and device longevity.

Purpose of the Study:

  • To evaluate the safety and efficacy of transvenous permanent pacemaker implantation in children weighing 15 kg or less.
  • To assess the complication rates and long-term outcomes in this pediatric population.

Main Methods:

  • A retrospective analysis of 24 children (≤15 kg) who underwent transvenous pacemaker implantation.
  • Follow-up ranged from 2 months to over 6 years, with detailed recording of complications and reoperations.

Main Results:

  • Eight out of 24 children experienced complications, with six requiring reoperation.
  • Complications included lead fracture, infection, myocardial dysfunction, generator migration, battery depletion, and threshold rise.
  • A redundant ventricular lead loop in the atrium successfully unraveled in all cases.

Conclusions:

  • Transvenous pacemaker implantation in small children is feasible, though complication rates necessitate careful monitoring.
  • The novel redundant lead technique appears safe and effective for managing lead length in growing children.
  • Most children achieve normal growth and development with acceptable cosmetic outcomes post-implantation.