His Bundle Pacing via the Atrial Port of a Dual-Chamber ICD: Programming Solution for Undersensing
Thai Duy Vo1, Le Uyen Phuong Tran1,2, Thi Van Khanh Nguyen3
1Department of Arrhythmology, Cho Ray Hospital, Ho Chi Minh city, Vietnam.
Background:
His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing.
Case Summary:
A 67-year-old man with permanent atrial fibrillation, alternating rapid and slow ventricular rates despite medication, symptomatic pauses longer than 6s, and a dual chamber ICD (His lead in the RA port; DF4 lead in the RV port) presented with palpitations. ECG and 24-hour Holter demonstrated intermittent His-channel undersensing with pacing on the T-wave peak. Interrogation showed small His electrograms followed by a prominent, constant ventricular far-field signal on the atrial channel. We turned off far-field protection after V-sense, intentionally used the ventricular far-field signal for sensing of His lead, and disabled related discrimination algorithms.
Conclusion:
A dual-chamber ICD with a His-bundle pacing lead may be associated with complex programming challenges, and targeted reprogramming can avoid lead revision or reimplantation.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Catheterization II: Right Heart Catheterization
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Catheterization III: Left Heart Catheterization
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy II: Dilated Cardiomyopathy