Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Accuracy comparison of a 16 and 64 multidetector-row computed tomography scanner to image small high-density
Empar Rollano-Hijarrubia1, Rik Stokking, Wiro J Niessen
1Department of Radiology, Erasmus MC - University Medical Center Rotterdam, Rotterdam, The Netherlands. m.rollanohijarrubia@erasmusmc.nl
Objectives:
The accuracy in imaging small high-density structures is compared for 16 and 64 multidetector-row computed tomography (MDCT) scanners.
Materials And Methods:
Phantom experiments and different quantification methods are used to establish size measurement accuracy, object signal, and image noise, for both MDCT systems.
Results:
At similar scanning doses, image noise is larger ( approximately 55%) for the 64 MDCT compared with the 16 MDCT, leading to lower signal-to-noise ratios ( approximately 28% for objects <2 mm). Object spread in the xy-plane is similar for both systems; while it is reduced along the z-axis (by approximately 0.18 mm) for the 64 MDCT. Measurement accuracy of the 64 MDCT is not significantly higher (P < 0.05) compared with the 16 MDCT when using a relative-threshold corresponding to 50% of the object maximum attenuation value. However, when using a fixed-threshold, interscanner and interprotocol measurement differences are statistically significant (eg, volume relative errors are reduced by approximately 17% on average for the 64 MDCT).
Conclusions:
Measurement accuracy of the 16 and 64 MDCT scanners is not significantly different when using a 50% relative threshold. However, image noise is significantly larger for the 64 MDCT. Compared with a fixed-threshold based method, the 50% relative-threshold strongly reduces interscanner and interprotocol measurement dependency and improves accuracy.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT