Background:

  • Current covered stents for pediatric use require large delivery sheaths and cannot be redilated after implantation.
  • These limitations pose challenges for minimally invasive procedures in children.

Purpose of the Study:

  • To develop and evaluate the clinical feasibility of a novel reexpandable covered stent for pediatric applications.
  • To assess the deliverability and redilation capability of the new stent in a preclinical model.

Main Methods:

  • An original Palmaz stent was modified with a polyurethane membrane, allowing up to 700% stretch.
  • The reexpandable covered stents were implanted in six mini piglets using a sheath slightly larger than standard.
  • Stent redilation was performed 28-70 days post-implantation, followed by macroscopic and microscopic examination.

Main Results:

  • The covered stents were successfully implanted in mini piglets via a relatively small sheath.
  • Redilation of the implanted stents was achieved, increasing diameter from 7.5 to 8.7 mm.
  • Histological analysis revealed good intimal coverage of the stent coating with minimal inflammation.

Conclusions:

  • The newly designed reexpandable covered stent demonstrates promising clinical feasibility for pediatric use.
  • This stent offers advantages in terms of smaller sheath delivery and post-implantation adjustability.
  • Further research is warranted to explore its long-term efficacy and safety in pediatric patients.