Superiority of synchrony of 256-slice cone beam computed tomography for acquiring pulsating objects. Comparison with
Purpose:
A prototype 256-slice cone beam computed tomography (CT) provides complete volumetric data within a single gantry rotation (1 s/rotation) with 0.5 mm slice-thickness.
Materials And Methods:
Calcified phantoms (200-400 HU) were attached to the balloon of a pulsating phantom and moved at a rate of 5-90/min. Acquisition was performed during one to-and-fro motion at each pulsation rate without electrocardiogram (ECG)-gating. Each period was divided into 10 phases, and compared to conventional multislice CT scanning without ECG-gating.
Results:
At 5-20/min, the configuration of calcified phantoms continued to the through-plane without gaps. At 60/min, duplicated calcified phantoms at end-systole and end-diastole were observed without motion. At 90/min, motion could be observed without gaps but was more blurred, and total calcified volume, Agatston scores, mean and max CT values of three phantoms were almost equal compared with those at static state. However, at 60/min, total calcified volume, scores, mean and max CT values of three phantoms were decreased to 64%, 37%, 80% and 56%, respectively, compared with those at static state. In multislice CT, even at lower rates, there were gaps in the through-plane. At 60/min, total calcified volume, scores, mean and max CT values of three phantoms were decreased to only 8%, 3%, 79% and 53%, respectively, compared with static state.
Conclusion:
This new prototype's unique character (synchrony) enables the acquisition of pulsating objects. These can be acquired without gaps in the through-plane even in the absence of ECG-gating. However, its present temporal resolution only permits accurate quantitative evaluation of calcium up to 20/min.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT