Sound speed correction in ultrasound imaging
David Napolitano1, Ching-Hua Chou, Glen McLaughlin
1Zonare Medical Systems, 1061 Terra Bella Ave., Mountain View, CA 94043, USA. djn@zonare.com <djn@zonare.com>
Abstract:
A constant sound speed of 1.54 mm/micros is generally used by ultrasound imaging systems for delay and timing. However, the body's sound speed in-homogeneity can lead to defocusing and increased clutter. To provide an improvement using standard transducers, the sound speed used in delay and timing was computed using different sound speeds. We observed improvement in lateral resolution and clutter in phantom, OB, abdominal, and breast imaging. In OB and abdominal imaging using a 4 MHz curved array, 1.48 mm/micros provided higher image quality in many situations. In breast with an 8 MHz linear array, 1.44 mm/micros provided better images in some cases. To provide an automated way to determine and adjust the sound speed used by the imaging system, an algorithm was developed that determines the sound speed that produces the best overall lateral image quality by analyzing the spatial frequency content in a single B-mode frame of channel data using images reconstructed using various trial sound speeds. The metric produced correlates well with the observed best lateral image quality.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Deriving the Speed of Sound in a Liquid
The speed of sound in fluids can be derived by considering a mechanical wave propagating...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies II: Ultrasonography
Speed of Sound in Solids and Liquids
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):