"Facial" and standard concentric needle electrodes are not interchangeable.
P E Barkhaus1, M M Roberts, S D Nandedkar
1Milwaukee Veterans Administration Medical Center/Medical College of Wisconsin, 53295, USA.
Different needle electrodes yield varying motor unit action potential (MUAP) amplitudes. Standard concentric needles (SCN) and facial concentric needles (FCN) are not interchangeable for EMG amplitude assessment.
Area of Science:
- Neurology
- Electrophysiology
- Biomedical Engineering
Background:
- Accurate recording of motor unit action potentials (MUAPs) is crucial for diagnosing neuromuscular disorders.
- Needle electrode selection can influence electromyography (EMG) signal characteristics, including amplitude and duration.
- Understanding these differences is vital for consistent and reliable EMG assessments.
Purpose of the Study:
- To compare the performance of standard concentric needle (SCN), facial concentric needle (FCN), and monopolar needle (MN) electrodes in recording MUAPs.
- To evaluate the impact of different needle types on MUAP amplitude and duration in specific muscles.
- To determine the interchangeability of these electrodes for EMG amplitude analysis.
Main Methods:
- MUAPs were recorded from the biceps brachii and first dorsal interosseous manus muscles of 7 healthy subjects.
- Three needle types were used: SCN, FCN, and MN.
- Analysis focused on MUAP duration and amplitude across the different electrode types.
Main Results:
- MUAP durations were comparable across all electrode types for each muscle.
- The FCN electrode recorded significantly higher amplitude MUAPs more frequently than the SCN electrode.
- While FCN and MN showed similar patterns of amplitude and duration changes, their uptake areas differed.
Conclusions:
- The FCN and SCN electrodes are not interchangeable for assessing EMG signal amplitude due to amplitude recording differences.
- Electrode type influences MUAP amplitude, necessitating careful selection for diagnostic accuracy.
- Different needle electrodes possess distinct signal acquisition characteristics impacting EMG interpretation.
More Related Videos
08:15Fabrication of Fine Electrodes on the Tip of Hypodermic Needle Using Photoresist Spray Coating and Flexible Photomask for Biomedical Applications
Published on: November 28, 2017
04:27Using Facial Electromyography to Assess Facial Muscle Reactions to Experienced and Observed Affective Touch in Humans
Published on: March 15, 2019
Related Concept Videos
Types of Reversible Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...