Primary Outcome Assessment in a Pig Model of Acute Myocardial Infarction
Published on: October 14, 2016
A pilot assessment of the FloTrac cardiac output monitoring system
Helen Ingrid Opdam1, Li Wan, Rinaldo Bellomo
1Department of Intensive Care, Austin Hospital, VIC 3084, Heidelberg, Australia. helen.opdam@austin.org.au
Objective:
To compare measurement of cardiac output (CO) by means of the FloTrac CO monitor with the pulmonary artery catheter (PAC).
Design:
Prospective observational study.
Setting:
Intensive care unit of a tertiary hospital.
Patients:
Six post-operative cardiac surgery patients with existing arterial cannulas and PACs.
Interventions:
Attachment of the FloTrac CO monitor and transducer to an existing arterial cannula. Simultaneous measurements of CO, indexed to body surface area (cardiac index, CI) by the FloTrac CO monitor and by either a bolus thermodilution or continuous CO PAC. Statistical analysis of observations.
Measurements And Results:
We performed CO measurements in six patients every 1-4 h after cardiac surgery. Comparison of all measurements showed a limited correlation for CI with the two devices (r (2)=0.1218, bias=0.21, 95% limits of agreement -0.81, 1.23). CI measurements obtained with the intermittent bolus PAC had better correlation with the FloTrac CI values (r (2) = 0.2693, bias=-0.0057, 95% limits of agreement -1.2042, 1.1929) than did those obtained with the continuous CO PAC (r (2)=0.0557, bias=0.2436, 95% limits of agreement -0.7350, 1.2222). When analysed according to heart rhythm, CI values measured during atrial pacing showed the best correlation (r (2)=0.377, bias=-0.0244, 95% limits of agreement -0.5226, 0.5714).
Conclusions:
CO measurements obtained using the FloTrac CO monitor show a limited correlation with those acquired using the PAC, relatively wide limits of agreement but no clear bias. Further evaluation is required before this device can be recommended for use in the clinical setting.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...