A packaged solution to the problems of testing arrhythmia control devices at implant
1Intermedics, Inc., Freeport, Texas 77515.
Pacing and Clinical Electrophysiology : PACE
|February 1, 1991
Summary
A novel packaging system simplifies implantable arrhythmia control device (ACD) testing. The ACD, within its sterile package, acts as the defibrillation test unit, reducing equipment needs and ensuring post-implantation function.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Implantable Electronic Devices
Background:
- Composite implantable arrhythmia control devices (ACDs) integrate multiple life-saving functions like pacing, cardioversion, and defibrillation.
- Current implant testing for ACDs is complex, time-consuming, and requires extensive, costly equipment.
- Improper testing can lead to critical device failure, such as failure to defibrillate.
Purpose of the Study:
- To introduce a unique packaging system for ACDs designed to streamline implant testing.
- To enable the ACD to function as its own defibrillation test unit during the implant procedure.
- To reduce the complexity and cost associated with ACD implant testing.
Main Methods:
- Development of a specialized sterile packaging system for the ACD.
- Integration of defibrillation testing capabilities directly into the ACD's packaging.
- Utilizing the ACD programmer and optional switch box for simplified testing procedures.
Main Results:
- The proposed packaging system allows the ACD to be tested while still in its sterile package.
- Significantly reduces the required support equipment to only the ACD, its programmer, and an optional switch box.
- Simplifies the implant testing procedure, making it quicker and easier to manage.
Conclusions:
- The novel packaging system enhances the efficiency and reliability of implantable arrhythmia control device testing.
- Physicians gain increased confidence in the ACD's correct function post-implantation due to testing with the actual implanted device.
- This innovation addresses the challenges of complex ACD testing, improving patient safety and procedural workflow.