Related Experiment Video
Updated: Jul 19, 2026

Synthesis and Operation of Fluorescent-core Microcavities for Refractometric Sensing
Published on: March 13, 2013
Quantitative reconstruction of refractive index distribution and imaging of glucose concentration by using diffusing
Xiaoping Liang1, Qizhi Zhang, Huabei Jiang
1J. Crayton Pruitt Family Department of Biomedical Engineering, University of Florida, Gainesville 32611, USA. xliang@bme.ufl.edu
Abstract:
We show that a two-step reconstruction method can be adapted to improve the quantitative accuracy of the refractive index reconstruction in phase-contrast diffuse optical tomography (PCDOT). We also describe the possibility of imaging tissue glucose concentration with PCDOT. In this two-step method, we first use our existing finite-element reconstruction algorithm to recover the position and shape of a target. We then use the position and size of the target as a priori information to reconstruct a single value of the refractive index within the target and background regions using a region reconstruction method. Due to the extremely low contrast available in the refractive index reconstruction, we incorporate a data normalization scheme into the two-step reconstruction to combat the associated low signal-to-noise ratio. Through a series of phantom experiments we find that this two-step reconstruction method can considerably improve the quantitative accuracy of the refractive index reconstruction. The results show that the relative error of the reconstructed refractive index is reduced from 20% to within 1.5%. We also demonstrate the possibility of PCDOT for recovering glucose concentration using these phantom experiments.