Related Experiment Video
Updated: Jul 19, 2026

Characterization of Proteins by Size-Exclusion Chromatography Coupled to Multi-Angle Light Scattering (SEC-MALS)
Published on: June 20, 2019
How much does band broadening adulterate results deduced from chromatograms measured by size-exclusion chromatography
Irene Schnöll-Bitai1, Christoph Mader
1Institute of Physical Chemistry, University of Vienna, Währinger Strasse 42, A-1090 Vienna, Austria. Irene.Schnoell-Bitai@univie.ac.at
Abstract:
A number of polystyrene samples and standards were measured with several column combinations which differed in their extent of band broadening, sigma(BB). Comparison of the derived chain length distributions showed in some cases good agreement even despite strong distinctions in the determined sigma(BB) values. The calculated number and weight averages of the samples were almost identical in most cases for the column combinations used. Furthermore, the points of inflection of the standards and of multimodale distributions composed of narrow (Poisson like) peaks in microemulsion were examined. When conferred with the theoretical values which were calculated for assumed Poisson distributions the respective deviations from the "true" ones were as high as 10% (almost reaching 20% in unfavorable cases). Simple (correction) procedures were tested in order to obtain the actual average values as well as the effective points of inflection of narrow distributions.
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Chromatographic Methods: Terminology
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...