Stabilization of a channeled spectropolarimeter by self-calibration
Atsushi Taniguchi1, Kazuhiko Oka, Hiroshi Okabe
1Division of Applied Physics, Graduate School of Engineering, Hokkaido University, Sapporo 060-8628, Japan.
Abstract:
A novel method to stabilize a channeled spectropolarimeter is described. The fluctuating retardations of the high-order retarders used in the spectropolarimeter are calibrated in parallel to the measurement of the wavenumber-dependent state of polarization (SOP) of light. Both the calibration of the retarders and the measurement of the SOP can be made simultaneously using a single light to be measured, and hence the resultant wavenumber-dependent SOP is almost immune to fluctuation of the retardations. The effectiveness of this method is experimentally demonstrated with retardation fluctuations induced by a temperature change up to 40 degrees C.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Pole and System Stability
Simple poles are unique roots of the denominator polynomial. Each simple pole corresponds to a distinct solution to the system's characteristic equation, typically resulting in exponential decay terms in the system's response.
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...